Data calibration

What is a calibration?

Calibrations

Geochemical analysis often require calibrating a machine. A calibration is a method that compares

To perform a calibration we usually need to measure a set of known samples (standards) in a machine.

 

Our standards will be a group of samples (usually artificial) containing no analyte (e.g a tube with water) or a known value of the analyte (e.g. a tube with liquid containing 10.2 parts per million of iron). Standards containing no analyte are usually called blanks, and the known values are often called nominal data. Excepting the blanks, the nominal values of the standards are expected to have an associated uncertainty (e.g. [Fe]=10.2±0.1 ppm).

Also, the machine we are using will be probably repeating the measurement on the same sample for 3 times or more. Therefore, the measured data will also have an associated uncertainty, usually the standard deviation of the measurements from the same sample.

Calibration tools

Plotting data with x and y errors

The first thing we should do to start calibrating any data is compare it graphically.

Create the script my_first_calibration.m and copy the following ICP calibration data:

    % Nominal concentrations of iron in some standards
    STDSnominal=[0,0.98,4.56,10.78,19.34,0,1.05,5.1,9.94,18.95];
    % associated uncertainties
    STDSnominal_uncert=[0,0.02,0.06,0.11,0.19,0,0.02,0.06,0.10,0.19];
    % ICP measured values of standards in counts per second (cps)
    STDScps=[425,1724,7443,15221,30973,146,1832,7378,15124,27701];
    % associated uncertainties
    STDScps_uncert=[214,140,377,329,381,249,311,280,1129,1140];

 

The simplest way of representing these data with uncertainties is:

    x=STDSnominal;
    dx=STDSnominal_uncert;
    y=STDScps;
    dy=STDScps_uncert;

    figure
    hold on
    for n=1:length(x)
      plot([x(n),x(n)],[y(n)-dy(n),y(n)+dy(n)],'-b')
      plot([x(n)+dx(n),x(n)-dx(n)],[y(n),y(n)],'-b')
    end
    plot(x,y,'.b')

 

However, when we have both x and y uncertainties, the error bars do not fully represent the probability distribution of the data in the 2-D space. Assuming that σx and σy are independent (no coviariance), the probability distribution at a point ( xi , yi ) can be defined by its Χ² value respect our data x, y, σx and σy:

image

Assuming that all the points with $\chi^2=1$ are at the one-sigma confidence level boundary, we could solve the previous equation as:

image

image

being τ between 0 and 2*π (note that image is always 1).  

We can use this property to draw the ellipses corresponding to our data within uncertainties:

    figure
    hold on
    for n=1:length(x)
      theta=linspace(0,2*pi,100);
      xi=x(n)+dx(n)*cos(theta);
      yi=y(n)+dy(n)*sin(theta);
      plot(xi,yi,'-b') % plot ellipse
    end
    plot(x,y,'+b') % plot central point

Note that ellipsis from samples with no uncertainty in one of the axis (e.g. blanks) look exactly as error bars.

Linear regression

A line is the simplest way of relating 2 sets of data (e.g. known concentrations and signals given by a machine). One of the most used methods to fit a line to our dataset is the “least-squares” regression. This method minimizes the square of the distances between the line and our data. Fortunately, there is a direct solution to solve this problem. The general formulas to fit a line y=a*x+b to n data by least-squares are:

image

 

In MATLAB/Octave, we can create the function leastsquares.m as:

    function [ myfit ] = leastsquares( x,y )
      a=(length(x)*sum(x.*y)-sum(x)*sum(y))/(length(x)*sum(x.^2)-sum(x)^2);
      b=(sum(x.^2)*sum(y)-sum(x)*sum(x.*y))/(length(x)*sum(x.^2)-sum(x)^2);
      myfit = @(x) a*x+b;
    end

and the average error of the data calibrated using leastsquares will be:

    myfiterror= mean(abs(y-myfit(x))

 

Exercises:

Interpolation and smoothing

When we have a curve defined as y=f(x) we might be interested in getting the xi value corresponding to a yi. This is the case of the function myfit(x), where x represent concentrations and y are signals obtained by ICP. The function leastsquares is a line and it would not be difficult to calculte the inverse function mathematically:

image

However, we often need to fill the gaps from incomplete datasets. For example, the file gistemp.csv (here) contains an estimate of global surface temperature change every 5 years (GISTEMP data).

Create a new script,load the data and plot it:

    gistempdata=csvread('gistemp.csv',1,0);
    years=gistempdata(:,1);
    temp=gistempdata(:,2);

    figure
    hold on
    plot(years,temp,'*k')

 

If we want to estimate the global surface temperature change every year, we need to interpolate the data. To do so, we can use the built-in function interp1. By default interp1(x,y,x0) will return the linear interpolation of the x,y dataset at x0. Try:

    myyears=min(years):1:max(years);
    mytemp=interp1(years,temp,myyears);
    plot(myyears,mytemp,'.-r')

To avoid errors, the x values in interp1(x,y,x0) should be sorted and not repeated. If your data is not sorted, you can use sort the data using sort: [x2,order]=sort(x); y2=y(order);

The linear interpolation is the default method used by interp1, so interp1(x,y,x0) is equivalent to interp1(x,y,x0,’linear’). But we can use other methods, such as ’spline’, or ’nearest’ to interpolate our data. To see the differences, try:

    myyears=min(years):1:max(years);
    mytemp=interp1(years,temp,myyears,'spline');
    plot(myyears,mytemp,'.-b')

Apart of the method, we can ask interp1 to also extrapolate data by adding ’extrap’ after the method. If you want to know more about interp1, type help interp1.

Exercise: use extrapolation to predict the global surface temperature change during the next century.

 

In other cases, instead of increasing the resolution of our data, we might be interested in smoothing it (e.g. to remove high frequency noise). For example:

     smoothingtime=50;
     yearssmooth=1900:10:2000;
     for n=1:length(yearssmooth)
        % select data around the year yearssmooth(n)
        selecteddata=(abs(yearssmooth(n)-years)<smoothingtime/2);
        tempsmooth(n)=mean(temp(selecteddata));
     end
     plot(yearssmooth,tempsmooth,'-m')

This method is called “moving average”.

Exercise: extrapolate the smoothed temperatures to predict the global surface temperature change during the next century.

Spectrometry data

Blank Equivalent Concentration

In spectrometry,the Blank Equivalent Concentration (BEC) is defined as the concentration that would correspond to the signal of the blank. It is usually determined by the following formula:

image

being I the signals measured, usually in counts per second (cps), and C the nominal concentration. Considering that we are going to be working with datasets involving several standards, we can define the BEC graphically as the negative of the x-intercept of our calibration line.

image

Graphical calculation of the Blank Equivalent Concentration (BEC) according to the calibration represented by the red line.

 

The BEC gives us an idea of how the background level of our machine compares with the measurements of our standards. We could think that a low BEC value implies that our measurements are going to be more precise. However, the precision of the measurements will depend on the stability of the background rather than the background value.

image

Two blank measurements compared. The measurements of the analyte #1 in the blank show a lower background, but the background of the analyte #2 measurements is more precise.

The stability of the background is what defines the precision of our measurements, rather then the background value. Similarly, we should calculate the variability of our BEC to get an idea of the noise of our measurements in concentration units. There are different ways of calculating the BEC “noise” (σBEC). We could just calculate it based on the scatter of our blank data. However, this would not reflect the scatter of our standards. The next figure shows the σBEC calculated from the uncertainty of our calibration.

Exercise: recycle the code generated before (calculation of myfiterror) to calculate the *uncertainty of the BEC corresponding to the calibration data.* Tip: you can use maths to get the inverse of myfit or use interp1.

image

Graphical calculation of the Blank Equivalent Concentration (BEC) and its uncertainty σBEC.

Limits of detection and quantification

The Limit of Detection (LOD) or Detection Limit is defined as the smallest measurable concentration. It is the concentration of a theoretical sample that will produce a signal strong enough to be distinguishable from the background noise. Assuming that this signal is going to have a noise similar to the background, the difference between the signals from the sample and the blank should be bigger than two times the noise. That is why the LOD is numerically defined as LOD=3noise. This is often calculated by calibrating the concentration corresponding to Iblank+3σIblank, being σIblank the standard deviation of the intensities measured on blank samples.

 

This approach assumes that our theoretical smallest measurable samples will produce a signal as scattered as our blank. However, sometimes, the measures of our samples are more similar to standards than to blanks (e.g. due to matrix effects). This is why considering LOD=3*σBEC would be a more conservative way of calculating our LOD.

Likewise, the Limit of Quantification (LOQ) is usually defined as 10 times the blank noise, so the uncertainty associated with the lowest sample that can produce quantitative data is $\sim 10$%. As for the LOD, we can calculate the LOQ using the BEC uncertainty: LOQ=10*σBEC.

Calibrating data

Once we have calculated the algorithms that relate the ICP signal with concentrations and concentration uncertainties, we are ready to calibrate the signals from our “unknown samples” in our script my_first_calibration.m:

    % ICP measured values of unknowns in counts per second (cps)
    SAMPLEScps=[9782,28746,13471,5870,28173,30492,13739,3588,813,12805];
    % associated uncertainties
    SAMPLEScps_uncert=[181,1042,1214,76,2899,2532,809,243,275,716];

To make this easier, we can define our calibration line as a reference xcal and ycal:

    xcal=linspace(min(x),max(x),100);
    ycal=myfit(xcal);

and transform the signals yunk into concentrations xunk using interp1:

    yunk=SAMPLEScps; dyunk=SAMPLEScps_uncert;
    xunk=interp1(ycal,xcal,yunk,'linear','extrap')

 

As the calibration is a line, we can also transform the measurement uncertainties into concentrations using:

    measurementuncert=...
        ( interp1(ycal,xcal,yunk+dyunk,'linear','extrap')-...
          interp1(ycal,xcal,yunk-dyunk,'linear','extrap') )/2

The measurement uncertainty is the internal uncertainty of our data, which is the errors we should use to compare our samples between them. However, as usually we want to compare our data with data that has not been calibrated simultaneously (e.g. samples measured one month ago), we should also include the calibration uncertainty into the external uncertainty (dxunk):

    calibrationuncert=...
      ( interp1(ycal,xcal,yunk+myfiterror,'linear','extrap')-...
        interp1(ycal,xcal,yunk-myfiterror,'linear','extrap') )/2
    dxunk=sqrt(calibrationuncert.^2+measurementuncert.^2);

 

Including the graphical representation of the unknown data, the script my_first_calibration.m could be something similar to this:

    %% This is my first calibration script

    clear % clear all previous data
    close all hidden % close all figures

    % Nominal concentrations of iron in some standards
    STDSnominal=[0,0.98,4.56,10.78,19.34,0,1.05,5.1,9.94,18.95];
    % associated uncertainties
    STDSnominal_uncert=[0,0.02,0.06,0.11,0.19,0,0.02,0.06,0.10,0.19];
    % ICP measured values of standards in counts per second (cps)
    STDScps=[425,1724,7443,15221,30973,146,1832,7378,15124,27701];
    % associated uncertainties
    STDScps_uncert=[214,140,377,329,381,249,311,280,1129,1140];
    x=STDSnominal; dx=STDSnominal_uncert;
    y=STDScps; dy=STDScps_uncert;

    % ICP measured values of unknowns in counts per second (cps)
    SAMPLEScps=[9782,28746,13471,5870,28173,30492,13739,3588,813,12805];
    % associated uncertainties
    SAMPLEScps_uncert=[181,1042,1214,76,2899,2532,809,243,275,716];
    yunk=SAMPLEScps; dyunk=SAMPLEScps_uncert;

    %% Calibration and uncertainty
    myfit = leastsquares(x,y);
    myfiterror= mean(abs(y-myfit(x)));

    %% Calibration line
    xcal=linspace(min(x),max(x),100);
    ycal=myfit(xcal);
    ycalerror=myfiterror;

    %% BEC, LOD, LOQ
    bec=-interp1(ycal,xcal,0,'linear','extrap');
    dbec=interp1(ycal,xcal,myfiterror,'linear','extrap')+bec;
    LOD=3*dbec;
    LOQ=10*dbec;

    %% Calibrate unknowns
    xunk=interp1(ycal,xcal,yunk,'linear','extrap');
    calibrationuncert=...
      interp1(ycal,xcal,yunk+myfiterror,'linear','extrap')-xunk;
    measurementuncert=...
      interp1(ycal,xcal,yunk+dyunk,'linear','extrap')-xunk;
    dxunk=sqrt(calibrationuncert.^2+measurementuncert.^2);

    %% Start a figure
    figure
    hold on

    % plot the unknowns with error-bars
    for n=1:length(xunk)
      plot(xunk(n),yunk(n),'.k')
      plot([xunk(n)-dxunk(n),xunk(n)+dxunk(n)],[yunk(n),yunk(n)],'-k')
      plot([xunk(n),xunk(n)],[yunk(n)-dyunk(n),yunk(n)+dyunk(n)],'-k')
    end

    % plot the standards with ellipsis
    for n=1:length(x)
      theta=linspace(0,2*pi,100);
      xi=x(n)+dx(n)*cos(theta);
      yi=y(n)+dy(n)*sin(theta);
      plot(xi,yi,'-b')
    end
    plot(x,y,'.b')

    % plot the calibration
    plot(xcal,ycal,'-r')
    plot(xcal,ycal+myfiterror,'--r')
    plot(xcal,ycal-myfiterror,'--r')

    % put labels
    ylabel('Intensity (cps)')
    xlabel('Concentration (ppm)')

 

And the generated figure will be similar to this:

image

Example of figure output of a calibration. Blanks and standards are depicted in blue, unknowns in black and the calibration line within uncertainty in red.

 

An example of a full propagation of uncertainties is shown in my_second_calibration.m.

 

image

Another example of calibration. Here, the scatter and the uncertainties of the standards are fully propagated.

 

Finally, you can convert your script into a function that you can use in the future to calibrate your own data:

    function [ SAMPLESppm SAMPLESppm_uncert] = calibfunction(STDSppm,STDSppm_uncert,...
                                                             STDScps,STDScps_uncert,...
                                                             SAMPLEScps,SAMPLEScps_uncert)

        % Paste here some of the code used in my_first_calibration
        %  - Remember no to paste the input data (STDSppm,SAMPLEScps, etc.)
        %  - Do not paste the plotting code unless you need it!

    end

Reporting data with uncertainties

At the end of our scripts we will probably want to include an easy way of exporting the numerical results we obtained, like the concentrations and uncertainties xunk and dxunk. The simplest way is by printing what we want in the command window using the built in function disp. As we want to mix our numerical parameters with strings, we will have to use num2str to transform our numbers into strings.

 

    clc % clear the command window
    disp(['Blank equivalent concentration: ' num2str(bec) ' ppm'])
    disp(['Limit of detection: ' num2str(LOD) ' ppm'])
    disp(['Limit of quantification: ' num2str(LOQ) ' ppm'])
    disp(['Concentrations and uncertainties:'])
    for n=1:length(xunk)
      disp([num2str(xunk(n)) ' +/- ' num2str(dxunk(n))])
    end
111

*If we want to be able to paste our copy and paste our data in Excel, we
can replace `’±’` by the tab character `char(9)`.*

 

A common mistake when reporting data with errors is using more digits
than the significant figures. For example, the last numbers of 6.3976
from the data 6.3976±0.46537 are meaningless. A couple of
significant figures in the uncertainty is usually enough, so in our
report we should just write 6.40±0.47, as this distribution is
identical to 6.3976±0.46537. We can use `round` to trim useless
digits from our `mu±sigma` data:

```Matlab
    decimalpositions=1-floor(log10(sigma));
    newmu=round(mu,decimalpositions);
    newsigma=round(sigma,decimalpositions);

 

Exercise: Add the “reporting code” in my_first_calibration.m to:

Exercise: ICP-OES data calibration

ICP-OES data calibration

The file ICPdata_GU20171012.csv (here) contains real raw ICP-OES data exported from the ICP machine (one line per analyte in chronological order) and an extra column with the nominal concentrations of the standards in ppm.

Write a script that reduces the ICP-OES data and report calibrated concentrations for each analyte.

image

ICP-OES at SUERC.

Tips

Some useful functions that you might need:

    for this_analyte=unique(analytename)’
        selectdata=strcmp(this_analyte, analytename);

        % ...and work with `selectdata` in here...

    end.